Chlamydia Trachomatis, Neisseria Gonorrhoeae sareng Trichomonas vaginalis
Ngaran Produk
HWTS-UR041 Chlamydia Trachomatis, Neisseria Gonorrhoeae sareng Trichomonas vaginalis Kit Deteksi Asam Nukléat (Fluoresensi PCR)
Épidemiologi
Chlamydia trachomatis (CT) nyaéta jenis mikroorganisme prokariot anu parasit pisan dina sél eukariot.Chlamydia trachomatis dibagi kana serotypes AK dumasar kana metode serotype.Inféksi saluran urogenital lolobana disababkeun ku trachoma varian biologis DK serotypes, jeung lalaki anu lolobana manifested salaku urethritis, nu bisa lega tanpa perlakuan, tapi lolobana aranjeunna jadi kronis, périodik aggravated, sarta bisa digabungkeun jeung epididymitis, proctitis, jsb.
Saluran
FAM | Chlamydia trachomatis |
ROX | Neisseria gonorrhoeae |
CY5 | Vaginitis trichomonal |
VIC/HEX | Kontrol internal |
Parameter Téknis
Panyimpenan | -18 ℃ |
Kahirupan rak | 12 bulan |
Jenis spésimén | swab cervical awéwé,swab heunceut awéwé,Swab uretra jalu |
Ct | ≤38 |
CV | <5% |
LoD | 400Salinan/mL |
Spésifikasi | Henteu aya réaktivitas silang sareng patogén inféksi STD sanés di luar rentang deteksi kit uji, sapertos Treponema pallidum, Mycoplasma hominis, Mycoplasma genitalium, Herpes simplex virus type 1, Herpes simplex virus type 2, Candida albicans, jsb. |
Instrumén lumaku | Applied Biosystems 7500 Real-Time PCR System Applied Biosystems 7500 gancang Real-Time PCR Systems QuantStudio®5 Sistem PCR Real-Time SLAN-96P Real-Time PCR Systems LightCycler®480 Sistim PCR Real-Time LineGene 9600 Ditambah Real-Time Sistim deteksi PCR MA-6000 Real-Time kuantitatif termal Cycler BioRad CFX96 Real-Time PCR System BioRad CFX Opus 96 Real-Time Sistim PCR |
Aliran Gawé
Pilihan 1.
Reagen ékstraksi anu disarankeun: Pipet 1mL sampel pikeun diuji kana 1.5mL tabung centrifuge bébas DNase/RNase, centrifuge dina 12000rpm salila 3 menit, piceun supernatant jeung jaga endapanana.Tambahkeun 200µL saline normal kana endapan pikeun resuspensi.Makro & Micro-Test Umum DNA/RNA Kit (HWTS-3019-50, HWTS-3019-32, HWTS-3019-48, HWTS-3019-96) (anu tiasa dianggo sareng Macro & Micro-Test Automatic Nucleic Acid Extractor (HWTS-3006C, HWTS-3006B)) ku Jiangsu Macro & Micro-Test Med-Tech Co., Ltd.. ékstraksi kudu dipigawé nurutkeun parentah pikeun pamakéan.Volume sampel anu diekstrak nyaéta 200µL, sareng volume élusi anu disarankeun nyaéta 80µL.
Pilihan 2.
Réagen ékstraksi anu disarankeun: Ekstraksi Asam Nukléat atanapi Kit Pemurnian (YDP302).Ekstraksi kudu dilaksanakeun sacara ketat nurutkeun parentah pikeun pamakéan.Volume élusi anu disarankeun nyaéta 80µL.